Main unit
■ 10.4"Colurful TFT+high-luminanceLCD Display, And keys
conditioning Means
■ Double-End pressure-flow sensor, such as tidal volume,
minute volume
■ Respiratory rate and airway pressure, pressure waveform, oxygen concentration, suffocation canbe real-time detection of manual control and automatic control
■ For the closed system, half-closed system and half-open system
■ Application: Adult and Child
■ Alarms parameters: Upper and lower airway pressure alarm; Upper and lower
■ Tidal volume alarm; Lack of oxygen alarm; Apnea; Power outages

Flow Meter
■ O2/N2O/AIR six-tube flow meter
■ Linkage of O2 and N2O; keep oxygen concentration over 25%
Vaporizer
■ Two high precise vaporizers(Enflurane、Halothane、Isoflurane、Sevoflurane)
■ Automatic compensation of flow, temperature and pressure
■ Self-lock and interlock
■ Volume: ≥250ml
Loop
■ Integrated design, orderliness respiratory loop
■ Quick and easy replacement
■ Intellectual property rights
SPECIFICATION
|
|
|
| Pneumatically driven and electrically control | |
| Monitoring Parameters: | Tidal volume, the respiratory rate; I/E Ratio; The peak airway pressure; Airway pressure wavefrom |
| Working mode: | Manual control or Automatic control |
| Ventilation modes: | IPPV, SIGH(Deep breaths/per time affer breathing 60 times) |
| SIMV/f/2, (Optional: | PEEP(+0.5kPa~+2 kPa))、SIPPV |
| Tidal volume: | 0,20~1500mL |
| Respiratory frequency: | 2~99bpm |
| I:E:5:1~1:8 adjustable | |
| PTRIG: | ,-2~0kPa |
| High pressure limitation: | 2~6kPa |
| Low pressure limitation: | 0~2kPa |
Picture