-three step defibrillation process
-two botton operation
-extensive voice and visual prompts for the operator
-Biphasic energy output
-lock-out protection to prevent inadvertent defibrillation
-automatic self-test: daily battery condition test&monthly AED whole intersystem test
Output | Specification |
Energy Sequence (user configurable) |
Biphasic Truncated exponential 150, 150, 200J |
Charge Time from “Shock Advised |
10 sec. to 150J 15 sec. to 200 J |
Analysis Time |
9 sec. |
From start of analysis to shock ready: |
from power up:Less than 20 seconds after 6 shocks:Less than 25 seconds |
Audible Prompts
|
22 audible prompts
|
Visual Prompts
|
LED prompts
|
Controls
|
Two buttons - On/Off, Shock
|
Output Energy Accuracy
|
±15% into any impedance from 25 to 175Ω,20Ω to 200Ω |
Power | Unit operates on internal battery only |
External Electrical Connections | No external devices are attached to the unit |
Risk Current Category | Internally powered equipment with defibrillator-proof BF type patient applied part (as per definition of IEC 60601-1 standard) |
Part No | CR123A-4x2 |
Non-rechargeable | 12V d.c. 2.8Ah |
Capacity | 100 discharges at 200 Joules or 120 discharges at 150 Joules |
Shelf Life (25℃±15℃) | 5 years |
Dimensions | 303 x 216 x 89 mm |
Weight | 2.0 kg |
Operating Temperature | 0℃ to 40℃ |
Operating Humidity | Relative humidity between 30% and 95% (non-condensing) |
Altitude | -300 to 15,000 ft( -91m to 4573m) |
Pressure | 102.4kPa~57.2kPa(768mmHg~429mmHg) |
Storage Temperature (without battery) | -20℃ to 55℃ |
Storage Humidity (without battery) | Up to 93% (non-condensing) |