• abnner

Multi-parameter Patient Monitor uMR 900N

12.1” color TFT display, high resolution: 800×600dpi;

Separated para board: ECG, NIBP, SpO2 board;

Full touch-screen enables intuitive operation by clicking one the specific parameters or the waveform in real time, standard configuration includes the navigation knob.


Product Detail

Product Tags

Features &Benefits:

1.12.1” color TFT display, high resolution: 800×600dpi;
2.Separated para board: ECG, NIBP, SpO2 board;
3.Full touch-screen enables intuitive operation by clicking one the specific parameters or the waveform in real time, standard configuration includes the navigation knob.
4.Real time S-T segment analysis, and ARR analysis;
5.SpO2 Pitch Tone Variation and Drug Dose Calculation;
6.Parameter waveform and characters color selectable
7.Multi-display: Standard, Large font, Trend Coexis, OxyCRG dynamic
8.Standby mode for Energy saving and suspend monitoring;
9.7-lead ECG waveforms display in phase;
10.Capture dynamic waveforms;
11.Built-in rechargeable lithium battery, and battery volume checking,
12.AC power checking
13.Efficient resistance to interference of defibrillator and HF knife;
14.SINNOR F-6 SpO2 technology, VS NELLCOR accuracy;
15.Networking capacity and TCP/IP networking platform enable abundant future upgrade
16.Time delay switch-off function keep the monitor data restart monitor;
17.Complete and flexible mounting solution for transportation and in-patient monitoring

Standard parameters: 5-lead ECG, SpO2(F-5),NIBP, TEMP,RESP,PR,HR.
Optional: Touch Screen, Thermal Recorder, 3/12-lead ECG, 2/4-IBP,4-TEMP,2-SpO2, Nellcor/Masimo SpO2,BIS, Phaisein/ Respironics Multi-gas, ICG/CO, FHR, FM, TOCO,

Performance Specifications

Display:       

12.1”color TFT

Rolling and refreshing waveform display

Resolution: 800×600

Multi display selectable including:

Large-font display

Trend coexist display

OxyCRG dynamic view display.

Bed-to-bed view display

 

Trace:             

9 waveforms(7 ECG, 1 SPO2 and 1 RESP)

Sweep speed:12.5mm/s,25mm/s,50mm/s

 

Indicator:          

Power/battery indicator light

QRS beep and alarm sound

 

Battery:           

Rechargeable lead acid cell, 12v/4AH

Max.24 hour for charging, 4 hours for continue working

 

Trend:            

Parameter graphic and tabular trends:

5s/piece, 8 hours;

1min/piece,168 hours(24hour×7days)

5min/piece,1000 hours.

Storage:           

NIBP: 1000 groups

Alarm: 200 groups

Full disclosure waveforms: 3600S

SD card external storage

 

Alarm:        

User-adjustable High, Medium and Low limits 3-level

Audible and visual alarm

 

Networking:        

Connected to central monitoring system

TCP/IP netting platform

 

Recorder:         

Build-in, thermal array

Plethysmogram caveform:2 channels

Record mode: manual, on alarm, time-defined

Recording width: 50mm

Printing speed: 50mm/s

 

Recording type:    

Frozen waveform record

NIBP recall record

Trend table record

Alarm record

Fixed -time record

Standard parameter:

ECG:

Lead mode: 5 –lead(R,L,F,N,C)

Lead selection : I,II,III,avR,avL,avF,V

Waveform: 3 and 7 channel selectable

Gain selection: 0.5mm/mv,1mm/mv,2mm/mv

Sweep speed: 12.5mm/s;25mm/s;50mm/s

Heart rate range:
Adult: 15~300bpm;
Neonate/pediatric:15~350bpm

Accuracy: +1bpm or +1%,whichecer is greater

Resolution: 1bpm

Filter:
surgery mode:1~20Hz
monitor model:0.5~40Hz
Diagnostic mode:0.05~130Hz

Scaling signal: 1mv,+3%

Protection: withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation

Alarm range: 15~350bpm

 

S-T segment detection:

Measurement range:  2.0mV~+2.0mV                         

Alarm Range:  -2.0mV~ +2.0mV

Accuracy: -0.8mV ~+0.8Mv

Error: +0.02Mv

 

Arrhythmia analysis: YES

RESPIRATION

Method: Thoracic impedance

Measurement range: Adult:7~120rpm;

Neonate/Pediatric:7~150rpm

Apnea alarm: YES,10~40s

Resolution:  1rpm

Accuracy:+2rpm

TEMPERATURE

Compatible probe: YSI or CYF

Measurement range: 5~50℃

Resolution :  0.1℃

Accuracy:  +0.1℃

Refreshing time:  about 1

Average measuring time:  <10s

FM

Manual button marking,

automatic FM identifying function

EtCO2( Sidestream CO2)

Measurement range:   0~99mmHg

Accuracy:   +2mmHg (0~40mmHg)

Sampling Range:   100ml/min

Sampling rate accuracy:   15%

Respiration rate:   0~120rmp

Respiration accuracy:  
+2rmp (0~70rmp)
+5rmp (>70rmp)

Respiration time:   
<240msec (10% to 90%)
Delay time:  <2s

Recorder:     

Build-in, thermal array

Plethysmogram caveform:2 channels

Record mode: manual, on alarm, time-defined

Recording width: 50mm

Printing speed: 50mm/s

Recording type:  
Frozen waveform record
NIBP recall record
Trend table record
Alarm record
Fixed -time record

SPO2:

Measurement range: 0~100%

Resolution: 1%

Accuracy: +2%(70-100%);0-69% unspecified

Alarm range : 0~100%

 

Pulse Rate:                

range: 20~300bpm

Resolution: 1bpm

Error: +1bpm or +2%,whichever is greater

NIBP

Method: Digital Automatic oscillometric

Operation mode : Manual/Automatic/continuous

Auto measurement time: Adjustable(1~480min)

Measurement Unit: mmHg/Kpa selectable

Measurement types:Systolic, Diastolic, Mean  

 

Measurement range:

Range of Systolic pressure:
Adult: 40~270mmHg
Pediatric: 40~220mmHg
Neonate: 40~135mmHg

Range of mean pressure:      
Adult: 20~235mmHg
Pediatric: 20~165mmHg
Neonate: 20~110mmHg

Range of diastolic pressure:   
Adult: 10~215mmHg
Pediatric: 10~150mmHg
Neonate :10~100mmHg

Over-pressure protection: Double safety protection

Resolution: 1 mmHg

Alarm: Systolic, Diastolic, Mean

TOCO Measuring

Transducer:  External pressure transducer

Measurement range:  0~100 units

Resolution: 1rpm

Accuracy: ±2 rpm

FHR

Transducer:   Multi-crystal, Pulsed Doppler

Measurement range:  50~210 BPM

Working frequency:   1 MHz

Strength:   <5mW/cm2

Signal processing:  special DSP system and modern recognization.

Resolution:   1BPM

Accuracy:    ±1BPM

Alarm Range:  
High: 160,170,180,190 BPM,
Low: 100,110,120 BPM

IBP

Channel: 2 channels

Range: -50-300mmHg

Resolution: 1mmHg

Accuracy: ±4mmHg(±4%)

Unit: mmHg, Kpa

Converter Sensitivity: 5mV/V/mmHg

Transducer sites: ART/PA/CVP/LAP/RAP/ICP

EtCO2 (Mainstream CO2)

Method:   Infrared Spectrum

Range:   0.0-10% (0~76%)

Resolution:   1mmHg (0.1%)

Accuracy:   
<5%(±4.0 mmHg)
Or<10%(of Readings)

Miscellaneous

Operation Environment

Temperature:    Working 0~+40℃

Transportation and storage –20~+60℃

Humidity:       

working≤85%

Transportation and storage≤93%

Power:          AC 100-240, 50/60Hz

Patient range:    Neonate, pediatric, and adult patients

Safety:              

Safety level: Class I, type CF

Dimension and Weight

Dimension:    37*21*43CM 

G.W.:        5.5KS

Accessories supplies

(1) 5 lead ECG cable  

(2) 1 spo2 probes  

(3) 1 NIBP probe  

(4) 1 temp probe

(5)1 ground connecting liner  

(6) Chest electrode (10pcs/set)

dcthf (1)
dcthf (2)
dcthf (3)
dcthf (4)

FAQ

1. How much extra if it is Touch Screen?

If you need to add touch screen, extra cost is $60.9.

2. How long is the guarantee?

It guarantees 12 months.

3. Is the battery RECHARGEABLE?

Yes, it is built-in rechargeable lithium battery and battery volume checking.

4. What certificates do you have?

Our patient Monitor have CE and ISO certificates etc.

 5. Do you support OEM service?

Yes, we can support OEM/ODM

For example, printing the customer's LOGO on the device or display the customer's LOGO name when it is turned on.


  • Previous:
  • Next:

  • Write your message here and send it to us

    Related Proutcts

    • Hyperbaric Oxygen Chamber uDR H1
    • Wholesale Popular Blood Glucose Meter Suit DR-G-001
    • Upper Arm Blood Pressure Monitor uHEM 910
    • Lightweight Foldable Electric Wheelchair Bumblebee X1
    • 12.1 inch standard 6 multi parameter parameter monitor uMR N15
    • Portable Veterinary Color Doppler  Ultrasound vDult P5