Wall mounting vital signs patient monitor uMR C8+
Specification
Display:
7’’color TFT
Rolling and refreshing waveform display
Multi display selectable including:
Large-font display
Trend coexist display
OxyCRG dynamic view display.
Bed-to-bed view display
Trace:
7 waveforms (5 ECG, 1 SPO2 and 1 RESP)
Sweep speed: 6.25mm/s, 12.5mm/s,25mm/s,50mm/s
Indicator:
Power/battery indicator light
QRS beep and alarm sound
Battery:
Rechargeable lead acid cell, 12v/4AH
Max.24 hour for charging, 4 hours for continue working
Trend:
Parameter graphic and tabular trends:
5s/piece, 8 hours;
1min/piece,168 hours(24hour×7days)
5min/piece, 1000 hours.
Storage:
NIBP: 1000 groups
Alarm: 200 groups
Full disclosure waveforms: 3600S
Alarm:
User-adjustable High, Medium and Low limits 3-level Audible and visual alarm
Networking:
Connected to central monitoring system
TCP/IP netting platform
Recorder:
Build-in, thermal array
Plethysmogram caveform: 2 channels
Record mode: manual, on alarm, time-defined
Recording width: 50mm
Printing speed: 50mm/s
Recording type:
Frozen waveform record
NIBP recall record
Trend table record
Alarm record
Fixed -time record
| Display: | 7’’color TFT |
| Rolling and refreshing waveform display | |
| Multi display selectable including: | |
| Large-font display | |
| Trend coexist display | |
| OxyCRG dynamic view display. | |
| Bed-to-bed view display | |
| Trace: | 7 waveforms (5 ECG, 1 SPO2 and 1 RESP) |
| Sweep speed: 6.25mm/s, 12.5mm/s,25mm/s,50mm/s | |
| Indicator: | Power/battery indicator light |
| QRS beep and alarm sound | |
| Battery: | Rechargeable lead acid cell, 12v/4AH |
| Max.24 hour for charging, 4 hours for continue working | |
| Trend: | Parameter graphic and tabular trends: |
| 5s/piece, 8 hours; | |
| 1min/piece,168 hours(24hour×7days) | |
| 5min/piece, 1000 hours. | |
| Storage: | NIBP: 1000 groups |
| Alarm: 200 groups | |
| Full disclosure waveforms: 3600S | |
| Alarm: | User-adjustable High, Medium and Low limits 3-level Audible and visual alarm |
| Networking: | Connected to central monitoring system |
| TCP/IP netting platform | |
| Recorder: | Build-in, thermal array |
| Plethysmogram caveform: 2 channels | |
| Record mode: manual, on alarm, time-defined | |
| Recording width: 50mm | |
| Printing speed: 50mm/s | |
| Recording type: | Frozen waveform record |
| NIBP recall record | |
| Trend table record | |
| Alarm record | |
| Fixed -time record |
Standard Parameter:
ECG:
Lead mode:
5 –lead(R,L,F,N,C)
Lead selection :
I,II,III,avR,avL,avF,V
Waveform:
3 and 7 channel selectable
Gain selection:
0.5mm/mv,1mm/mv,2mm/mv
Sweep speed:
12.5mm/s;25mm/s;50mm/s
Heart rate range:
Adult: 15~300bpm; Neonate:/pediatric:15~350bpm
Accuracy:
+1bpm or +1%, whichecer is greater
Resolution:
1bpm
Filter:
1~20Hz
monitor model:
0.5~40Hz
Diagnostic mode:
0.05~130Hz
Scaling signal:
1mv,+3%
Protection:
withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation
Alarm range:
15~350bpm
S-T segment detection:
S-T segment detection:
Measurement range:
2.0mV~+2.0mV
Alarm Range:
-2.0mV~ +2.0mV
Accuracy:
-0.8mV ~+0.8Mv
Error:
+0.02Mv
Arrhythmia analysis:
YES
ECG:
| Lead mode: | 5 –lead(R,L,F,N,C) |
| Lead selection : | I,II,III,avR,avL,avF,V |
| Waveform: | 3 and 7 channel selectable |
| Gain selection: | 0.5mm/mv,1mm/mv,2mm/mv |
| Sweep speed: | 12.5mm/s;25mm/s;50mm/s |
| Heart rate range: | Adult: 15~300bpm; Neonate:/pediatric:15~350bpm |
| Accuracy: | +1bpm or +1%, whichecer is greater |
| Resolution: | 1bpm |
| Filter: | 1~20Hz |
| monitor model: | 0.5~40Hz |
| Diagnostic mode: | 0.05~130Hz |
| Scaling signal: | 1mv,+3% |
| Protection: | withstand 4000VAC/50 voltage isolation against electrosurgical interference and defibrillation |
| Alarm range: | 15~350bpm |
| S-T segment detection: | S-T segment detection: |
| Measurement range: | 2.0mV~+2.0mV |
| Alarm Range: | -2.0mV~ +2.0mV |
| Accuracy: | -0.8mV ~+0.8Mv |
| Error: | +0.02Mv |
| Arrhythmia analysis: | YES |
SPO2
Measurement range:
0~100%
Resolution:
1%
Accuracy:
.+2%(70-100%);0-69% unspecified
Alarm range :
0~100%
Pulse Rate:
range: 20~300bpm
Resolution:
1bpm
Error:
.+1bpm or +2%,whichever is greater
SPO2
| Measurement range: | 0~100% |
| Resolution: | 1% |
| Accuracy: | .+2%(70-100%);0-69% unspecified |
| Alarm range : | 0~100% |
| Pulse Rate: | range: 20~300bpm |
| Resolution: | 1bpm |
| Error: | .+1bpm or +2%,whichever is greater |
NIBP:
| Method: | Digital Automatic oscillometric |
| Operation mode : | Manual/Automatic/continuous |
| Auto measurement time: | Adjustable(1~480min) |
| Measurement Unit: | mmHg/Kpa selectable |
| Measurement types: | Systolic, Diastolic, Mean |
| Measurement range: | Measurement range: |
| Range of Systolic pressure: | Adult: 40~270mmHg |
| Pediatric: 40~220mmHg | |
| Neonate: 40~135mmHg | |
| Range of mean pressure: | Adult: 20~235mmHg |
| Pediatric: 20~165mmHg | |
| Neonate: 20~110mmHg | |
| Range of diastolic pressure: | 10~215mmHg |
| Pediatric: 10~150mmHg |
|
| Neonate :10~100mmHg | |
| Over-pressure protection: | Double safety protection |
| Resolution: | 1 mmHg |
| Alarm: | Systolic, Diastolic, Mean |
RESPIRATION
| Method: | Thoracic impedance |
| Measurement range: | Adult:7~120rpm; |
| Neonate/Pediatric:7~150rpm | |
| Apnea alarm: | YES,10~40s |
| Resolution: | 1rpm |
| Accuracy: | . +2rpm |
TEMPERATURE
| Compatible probe: | YSI or CYF |
| Measurement range: | 5~50℃ |
| Resolution : | 0.1℃ |
| Accuracy: | .+0.1℃ |
| Refreshing time: | about 1 |
| Average measuring time: | <10s |
Miscellaneous
Safety:
Safety level: Class I, type CF
Operation Environment
Temperature: Working 0~+40℃
Transportation and storage –20~+60℃
Humidity: working≤85%
Transportation and storage≤93%
Power: AC 100-240,50/60Hz
Patient range: Neonate, pediatric, adult patients
Accessories supplies
(1) 5 lead ECG cable
(2) 1 spo2 probes
(3) 1 NIBP probe
(4) 1 temp probe
(5)1 ground connecting liner
(6) Chest electrode (10pcs/set)
Dimension and Weight
Dimension: 31*19*41CM
G.W.: 3KGS