1: |
Summary of main specifications and system of cart type color Doppler ultrasound |
1.1 |
Trolley type all digital color Doppler ultrasonic mainframe |
1.2 |
Ultrasonic host operating system: Windows operating system |
1.3 |
Applications:Abdomen, obstetrics, gynecology, heart, urinary system, small organs, superficial, blood vessels, pediatrics, newborns, musculoskeletal |
1.4 |
Probes:Convex probe, Tran-vaginal probe, Linear probe, Micro-convex probe, Cardiac probe,4D Volumeprobe |
1.5 |
Applications and report:Abdominal,OB,GYN,Cardiac,Urinary,Small Parts,Superficial, Vascular, Pediatrics, Advanced measurement software packages, report software packages, case management software packages, etc. |
☆1.6 |
carotid artery intima measurement thickness (IMT) |
☆1.7 |
Automatic spectral envelope measurement |
1.8 |
Full digital transmission and reception of beam synthesizer |
1.9 |
Color Doppler imaging(C) |
1.10 |
Pulse Doppler Imaging(PW) |
☆1.12 |
Coherent Contrast imaging(CCI) |
1.12 |
Continuous wave Doppler imaging(CW) |
☆1.13 |
B/C/D Real-time three synchronous imaging |
☆1.14 |
Power Doppler imaging(PDI) |
☆1.15 |
Direct Power Doppler imaging(DPDI) |
1.16 |
M mode imaging |
☆1.17 |
Anatomic M mode imaging |
☆1.18 |
Color Doppler M mode imaging |
☆1.19 |
Elastography |
☆1.20 |
Tissue Doppler imaging(TDI) |
☆1.21 |
Strain rate imaging (SRI) |
1.22 |
Tissue harmonic imaging(THI) |
1.23 |
Fusion harmonic imaging(FHI) |
1.24 |
Speckle Reduce imaging(SRI) |
☆1.25 |
Panoramic imaging |
☆1.26 |
Deflection imaging |
☆1.27 |
Trapezoidal imaging |
1.28 |
Adaptive velocity optimization |
☆1.29 |
Free hand 3D |
1.30 |
Real time 3D imaging(3D/4D) |
1.31 |
DICOM3.0 |
1.32 |
Monitor:≥21.5 inch,high definition ultrasonic display |
1.33 |
≥13.3 inch touch screen |
1.34 |
Physical clipboard: save the image on the left side of the screen, which can be directly transferred or deleted. |
1.35 |
The system has the function of on-the-spot upgrade |
1.36 |
Presupposition: for different inspection of the viscera, preset the inspection conditions for the best image, reduce the adjustment of the operation, and the commonly used external adjustment and combination regulation. |
1.37 |
Probe interface: 4 |
1.38 |
Chinese and English System, Chinese and English input, optional |
1.39 |
Depth:≥360mm; |
1.40 |
Extended imaging |
2: |
Probes |
2.1 Convex probe |
Fundamental Frequency:2.0MHz/2.3MHz/2.5MHz/3.0MHz/3.5MHz/4.0MHz/4.6MHz/5.0MHz/5.4MHz,Harmonic Frequency:4.0MHz/4.6MHz/5.0MHz, |
2.2 Linear probe |
Fundamental Frequency:4.0MHz/4.6MHz/5.0MHz/6.0MHz/7.0MHz/8.0MHz/9.2MHz/10.0MHz/12.0MHz/13.3MHz,Harmonic Frequency:8.0MHz/9.2MHz/10.0MHz, |
2.3 Rectal probe |
Fundamental Frequency:4.0MHz/4.6MHz/5.0MHz/6.0MHz/7.0MHz/8.0MHz/9.2MHz/10.0MHz/12.0MHz/13.3MHz,Harmonic Frequency:8.0MHz/9.2MHz/10.0MHz, |
2.4 Micro-convex probe |
Fundamental Frequency:3.0MHz/3.5MHz/4.0MHz/5.0MHz/5.4MHz/6.0MHz/7.0MHz/8.0MHz,Harmonic Frequency:
6.0MHz/7.0MHz/8.0MHz, |
2.5 Cardiac probe |
Fundamental Frequency:1.7MHz/1.9MHz/2.1MHz/2.5MHz/3.0MHz/3.4MHz/3.8MHz/4.2MHz/5.0MHz,Harmonic Frequency:3.4MHz/3.8MHz/4.2MHz, |
2.6 4D Volume probe |
Fundamental Frequency:2.0MHz/2.5MHz/3.0MHz/3.3MHz/3.7MHz/4.0MHz/5.0MHz/6.0MHz,Harmonic Frequency:
4.0MHz/5.0MHz/6.0MHz, |
3: |
2D imaging mode |
3.1 |
Gain:0-100,Step 2 adjustable |
3.2 |
TGC:8 segment adjustable |
3.3 |
Maximum focus point:≥7, which can be moved throughout the whole process. |
3.4 |
Speckle reduction:0-5,5 level |
3.5 |
Space Synthesis:0-2,2 level(Liner probe: 3 level, cardiac probe:0) |
3.6 |
Dynamic:30-180,35 level,step 5 adjustable |
3.7 |
Line density:low、middle、high,3 level |
3.8 |
Frame correlation:0-4, 4 level |
3.9 |
Noise reduction:0-5,5 level |
3.10 |
Edge Enhancement:0-5,5 level |
3.11 |
Sound power:2-10, 9 level |
3.12 |
Grey scale:0-67, 67 level |
3.13 |
False color:0-67,67 level |
3.14 |
Image style:Soft-Comparison,2 level |
|
The screen has real-time display of voice power, probe frequency, dynamic range, pseudo color, gray scale and other 11 parameters can be adjusted |
4: |
Color Doppler Imaging Mode |
4.1 |
Blood gain:0-100,Step 2 |
4.2 |
Parameter display:Velocity、Variance |
4.3 |
B-Restrain(B/W restrain):0-7, 7 level |
4.4 |
Speed Through:0-8, 8 level |
4.5 |
Sampling number:6-24, 7 level |
4.6 |
Blood flow preferred:0-8, 8 level |
4.7 |
Filtering:1-6, 6 level |
4.8 |
Sound power:2-6, 4 level |
4.9 |
Noise reduction:0-4, 4 level |
4.10 |
Smooth treatment:0-4, 4 level |
4.11 |
Frame correlation:0-6, 6 level |
4.12 |
Chromatography(Blood flow graph):0-37, 37 level |
4.13 |
Line density:Low-Middle-High, 3 level |
4.14 |
Frequency:4 level adjustable |
4.15 |
Velocity:Minimum 0.4K,Maximum 40.5KConvex probe:0.4K-4.3K-38.5KLinear probe:0.4K-14.7K-39.0K
Trans-vaginal probe:0.4K-7.8K-39.7K Volume probe:0.4K-4.2K-34.8K Micro-convex probe:0.4K-10.3K-40.5K cardiac probe:0.4K-7.8K-39.7K |
|
PS:The frequency of the probe changes and the frequency value changes |
|
PS:Frame rate changes with speed |
5: |
Pulse wave Doppler(PW) |
1 |
Gain:0-100,Step 2 |
5.2 |
Spectrum envelope function: real time automatic spectrum envelope, manual spectrum envelope, and other modes. The system automatically analyses and displays various data such as PSV, EDV, RI, PI, S/D, ACC, HR and so on. Can wake up or close |
5.3 |
Sample volume:0.5mm~30mm |
5.4 |
Blood angel:-75—75 degree,Step 5 |
5.5 |
False color:0-67, 67 level |
5.6 |
Dynamic range:20-40, 4 level |
5.7 |
Filter:0-9, 9 level |
5.8 |
Smooth treatment:1-4, 4 level |
5.9 |
Sound power:2-5, 4 level |
5.10 |
Volume:0-100, 10 level,Step 10 |
5.11 |
Audio filtering:0-4, 4 level |
5.12 |
Base line:-1.0~1.0, |
5.13 |
Grey map:0-67, 67 level |
5.14 |
Scan velocity:100-500, 6 level |
5.15 |
PRF:Minimum 0.5K,Maximum 87.5KConvex probe:0.5K-4.3K-63.3KLinear probe:0.5K-14.5K-78.4K
Trans-vaginal probe:0.5K-8.1K-78.4K Volume probe:0.5K-4.2K-53.8K Micro-convex probe:0.5K-10.3K-81.1K cardiac probe:0.5K-4.3K-87.5K |
5.16 |
Frequency:4 level |
|
PS:The frequency of the probe changes and the PRF value changes |
|
PS:The frequency of the probe changes and the frequency value changes |
6: |
Continuous Wave Doppler(CW) |
6.1 |
Support probe:Cardiac probe |
6.2 |
Adjustment of B mode parameters is switchable |
6.3 |
Gain:0-100,Step 2 |
6.4 |
Sampling line position is adjustable |
6.5 |
PRF:0.9K~36.1K |
6.6 |
Baseline:-1.0~1.0 |
6.7 |
Blood angel:-75~75 degree |
6.8 |
Grey map:0-67 |
6.9 |
Scan velocity:100-300 |
6.10 |
False color:0-67 |
6.11 |
Dynamic range:20-40 |
6.12 |
Filtering:0-9,9 level |
6.13 |
Smooth treatment:1-4 |
6.14 |
Frequency:2.0MHz/2.3MHz/2.5MHz/3.0MHz,4 level adjustable |
6.15 |
Sound power:2-5 |
6.16 |
Volume:0-100 |
6.17 |
Audio Filtering:0-4 |
☆7: |
Anatomical M imaging |
7.1 |
Support probe:Convex probe, Linear probe,Cardiac probe |
7.2 |
Adjustment of B mode parameters is switchable |
7.3 |
Gain:0-100,Step 2 |
7.4 |
M Sampling line angel is adjustable |
7.5 |
M Sampling line length is adjustable |
7.6 |
Sampling line:3,Can be displayed or hidden separately |
☆8: |
Blood flow M mode(MC) |
8.1 |
Adjustment of B mode parameters is switchable |
8.2 |
Gain:0-100,Step2 |
8.3 |
MC Sampling line angel is adjustable |
8.4 |
MC Sampling line length is adjustable |
8.5 |
Frequency:4 level |
8.6 |
Sampling number:6-24 |
8.7 |
Speed through:0-8, 8 level |
8.8 |
Scan velocity:150-500 |
8.9 |
Frame correlation:0-6, 6 level |
8.10 |
Filtering:1-6,6 level |
8.11 |
Blood flow preferred:0-8,8 level |
8.12 |
Smooth treatment:0-4,4 level |
8.13 |
Map:0-37, 37 level |
☆9: |
Elastography |
9.1 |
Adjustment of B mode parameters is switchable |
9.2 |
Gain:0-100,Step 2 |
9.3 |
B/E,Double real-time display on the same screen |
9.4 |
Probe displacement curve display:Up/Down |
9.5 |
Pressure indicator bar display |
9.6 |
Frequency:8-9 level,Adjustable;According to the probe display |
9.7 |
Noise reduction:0-2, 2 level |
9.8 |
Frame correlation:0-3, 3 level |
9.9 |
Comparison:0-13, 13 level |
9.10 |
False color:0-3, 3 level |
9.11 |
Don’t support cardiac probe |
☆10: |
Tissue Doppler imaging(TDI) |
10.1 |
Support probe: Cardiac probe |
10.2 |
Adjustment of B mode parameters is switchable |
10.3 |
Gain:0-100,step 2 |
10.4 |
ROI area adjustable |
10.5 |
Sampling number:6-24 |
10.6 |
Velocity:0.4K-8.0K |
10.7 |
Frame correlation:0-6,6级 |
10.8 |
Tissue preferred:0-7, 7 level |
10.9 |
Frequency:2.0MHz/2.3MHz/2.5MHz/3.0MHz |
10.10 |
Support color reversal |
☆11: |
Strain rate imaging |
11.1 |
Support probe: Cardiac probe |
11.2 |
Adjustment of B mode parameters is switchable |
11.3 |
ROI area adjustable |
11.4 |
Gain:0-100,Step 2 |
11.5 |
Sampling number:6-24,6 level |
11.6 |
Axial average:1-4, 4 level |
11.7 |
Velocity:0.4K-8K |
11.8 |
Frame correlation:0-6, 6 level |
11.9 |
Tissue optimization:0-7,7 level |
☆12: |
Panoramic imaging |
12.1 |
Support probe:Linear probe |
12.2 |
Speckle Reduction:0-5, 5 level |
☆13: |
Deflection imaging |
13.1 |
Support probe:Linear probe |
13.2 |
Adjustment of B mode parameters is switchable |
13.3 |
Deflection angel:8 level |
13.4 |
Speckle reduction:0-5, 5 level |
13.5 |
Dynamic rate:30-180,Step 5 |
13.6 |
Line density:low-middle-high,3 level |
13.7 |
Frame Correlation:0-4, 4 level |
13.8 |
False color:0-67, 67 level |
13.9 |
Image style:Soft-Comparison,2 level |
13.10 |
Noise reduction:0-5, 5 level |
13.11 |
Edge Enhancement:0-5,5 level |
13.12 |
Sound power:2-10,8 level |
13.13 |
Grey map:0-67,67 level |
☆14: |
Trapezoidal imaging |
14.1 |
Probe support:linear probe |
14.2 |
Adjustment of B mode parameters is switchable |
14.3 |
Deflection angel:8 level |
14.4 |
Speckle reduction:0-5, 5 level |
14.5 |
Dynamic rate:30-180,Step 5 |
14.6 |
Line density:low-middle-high,3 level |
14.7 |
Frame Correlation:0-4, 4 level |
14.8 |
False color:0-67, 67 level |
14.9 |
Image style:Soft-Comparison,2 level |
14.10 |
Noise reduction:0-5, 5 level |
14.11 |
Edge Enhancement:0-5,5 level |
14.12 |
Sound power:2-10,8 level |
14.13 |
Grey map:0-67, 67 level |
14.14 |
Space Synthesis:0-2, 2 level |
☆15 |
Freehand 3D imaging |
15.1 |
Support probe:convex probe, linear probe |
15.2 |
Display model: 4 pictures |
15.3 |
Image Rotation X/Y/Z Axis |
15.4 |
Multi-slice Visibility |
16 |
Real-time 4D imaging |
16.1 |
Support probe: 4D volume probe |
16.2 |
Adjustment of B mode parameters is switchable |
16.3 |
Gain:0-100,Step 2 |
16.4 |
Display model:one image、two images、four images |
16.5 |
Image Rotation:X/Y/Z Axis |
16.6 |
Multi-slice Visibility |
16.7 |
Light&Shade inversion |
16.8 |
Smooth:0-4, 4 level |
16.9 |
Threshold level:0-129, Step 3 |
16.10 |
Transparency:1-509,Step 10 |
16.11 |
Render type:4 kinds,Surface、maximum、minimum、perspective |
17: |
Extended Imaging |
17.1 |
Gain:0-100,Step 2 |
17.2 |
TGC:8 segment adjustable |
17.3 |
Maximum focus point:≥7, which can be moved throughout the whole process. |
17.4 |
Speckle reduction:0-5,5 level |
17.5 |
Space Compound:0-2,2 level(Linear probe: 3 level,don’t support cardiac probe) |
17.6 |
Dynamic range:30-180,35 level,Step 5 |
17.7 |
Line density:Low、Middle、High,3 level |
17.8 |
Frame correlation:0-4,4 level |
17.9 |
Noise reduction:0-5,5 level |
17.10 |
Edge enhancement:0-5,5 level |
17.11 |
Sound power:2-10, 9 level |
17.12 |
Grey map:0-67, 67 level |
17.13 |
False color:0-67,67 level |
17.14 |
Image style:Soft-Comparison,2 level |
17.15 |
Extended level: Maximum 72 levelConvex probe:9 levelTrans-vaginal probe:72 level
Micro-convex probe:29 level Cardiac probe:40 level 4D Volume probe:17 level |
|
PS:The screen has real-time display of voice power, probe frequency, dynamic range, pseudo color, gray scale and other 11 parameters can be adjusted |
|
PS: When the probe scan range reaches the maximum, the space synthesized is 0. |
18: |
Measurement and analysis function: |
18.1 |
General measurement: Distance, area, ellipse, cross line, angle, distance ratio, volume, Volume (ellipse), area ratio, diameter, joint angle |
18.2 |
Cardiac:Automatic spectrum envelope、LV、Main Pulmonary artery diameter、RVEDd、RVEDs、LVM、LAV、HR、MVF、AO、AR、LVOT、TVF、Pulmonic valve、Pulmonary vein、RV、Doppler fetal heart sound、LVET、LVM、LVMI、AV |
18.3 |
Vascular: carotid intima (IMT), length stenosis ratio, area stenosis ratio, IMT (back wall), IMT (front wall) |
18.4 |
OB:Fetal routine、AFI、TW、GS、CRl、OFD、HL、ulna、NT、Fibula、Nbonel、Radial、Tibia |
18.5 |
GYN:uterus、cervix、corpus uteri/cervix uterus、left ovarian vein、right ovarian vein、dominant follicle、intima thickness |
18.6 |
Urology: prostate、residual urine、left kidney、right kidney、left suprarenal vein、right suprarenal、left testis、right testis、left seminal vesicle、right seminal vesicle |
18.7 |
Abdomen:liver、CHD、partal vein diameter、cholecyst、CBD、pancreas、spleen、Internal diameter of abdominal aorta、kidney |
18.8 |
Small parts:Thyroid |
18.9 |
Software package:Measurement package、Software package、Medical records management software package |
19: |
Graphic and text management system |
19.1 |
Host build in 2 hard disk(SSD 120+1T),Start fast and stable |
19.2 |
Movie playback:≥1200 frames |
19.3 |
Internal file information management system: can record patient number, name, check number, check date and so on, and can be searched and managed by numbering, checking number, name and so on. |
19.4 |
Type of report is 16 |
19.5 |
One key fast report graphic and text management |
20: |
Interface |
20.1 |
USB interface: 4 |
20.2 |
HDMI interface: 1 |
20.3 |
RJ-45 interface:1 |
20.4 |
Grounding wire interface:1 |
20.5 |
DVD RW:1 |
21: |
Configuration |
21.1 |
Trolley type full digital color Doppler ultrasound diagnostic system |
21.2 |
Probe: convex array probe (standard), linear probe (optional), Trans-vaginal probe (optional), cardiac probe (optional), 4D volume probe (optional) |
21.3 |
≥13 quick adjusting knobs |
22: |
Technology, after-sales service and other requirements |
22.1 |
After acceptance, the warranty is free for two years(Provide manufacturer warranty certificate) |
22.2 |
Manufacturer has ISO13485 certification and EU CE certification. |