-smart,compact,precise and reliable
-portable,only 12kg total
-Economical:Integrate hydraulic design user less reagents for testing, reduce the cost for users
-Fast&Convenient: fast test, 60t/h, English operation interface, optional touch screen
-simple operation and low consumption
-advanced design and stable performance
-accurate analysis
Function: automated WBC 3-part
Test parameters:
-21 parameters: WBC, Neu#, Lym#, Mon#, Eos#,Bas#,Neu%, Lym%, Mon%, Eos%, Bas%, RBC,HGB, HCT, MCV,MCH, MCHC, RDW-SD,RDW-CV, PLT, MPV, PDW, PCT.
-3 histograms for WBC, RBC, and PLT
Principal: electronic impedance, SFT method for hemoglobin
Throughout: 60 tests/hour
Analysis mode: optional venous blood
Sample volume: anticoagulant whole blood<9.6ul,capillary blood<9.6ul
Working model: automatic sampling, detecting automatic testing, furnished with dilutor, no need diluting outside. Automatic dispensing lyse and making second diluting automatically.
Block clearing: soak and washing back and forth , high pressure burning. Boast the function of monitoring incomplete blockage
Alarm: automated system diagnosis, self testing, fault alarm and reagent shortage alarm
Automated maintenance: automated washing during power one and power off,reducing the blockage rate
Standby protection: 24 hours switch on, automatic sleep and wake up function
Storage: large quantity test result storage , at least 40 thousand test results with
Control Mode: Quality control mode of L-J,X-B and X-r. west guard quality control regulation
Calibration mode: manual and automatic calibration available , fresh blood calibration available also
Reference value setting: many reference value range(children, adult and the aged). The appropriate reference value range could be automatically selected according to the inputted aga and gender
Discriminator function: Float and fix discriminator, manual adjust available
Display: TFT LCD(Optional touch screen)
Input Mode: standard mouse, optional bard code reader
Print Mode: internal high speed thermal printer ,optional connection with specified external printer, English printouts
Report Format: Many kinds of report format(not less than 8 kinds)
Reagents System: Open reagent system, no cyanogen reagents. Provide original reagents
Connaction port: USB, RS-232, network interface, LPT
Dimension: 345mm*245mm*330mm
Weight: 12KG