| Display |
| Screen Type: |
High-resolution display, TFT colour display |
| Screen Size: |
7 inches (17.7cm) diagonally |
| Sweep Speed: |
12.5/25/50mm/sec |
| Information: |
Heart Rate, Lead, Alarm On/Off, Delivered Energy |
| Resolution: |
800X480 |
| ECG |
| Patient Connection: |
3-lead ECG cable, or 5-lead ECG cable, paddles. |
| Lead Selection: |
Displayed on monitor, paddles, I, II, III,AVR, AVL, AVF, V. |
| ECG Size: |
2.5,5, 10, 20, 30 mm/mV display on monitor. |
| Heart Rate: |
20-350BPM. |
| Protection: |
With stand 4000vAc/50Hz voltage in isolation and work against electrosurgical knife and defibrillation |
| Defibrillator |
| Waveform: |
Biphasic |
| Defibrillator type: |
Manual External Synchronous/Asynchronous Defibrillation. |
| Charge Time: |
Less than 10 seconds to 360 Joules |
| Energy Display: |
Monitor display indicates both selected and delivered energy. |
| Charge Prompt Type: |
Voice and visual prompts. |
| Energy Selection: |
Selectable at 0, 3, 5, 7, 10, 20, 30, 50, 100, 200, 300, 360 joules |
| Standard Adult/Pediatric Paddles: |
Reusable external adult paddles (pediatric paddles integrated) |
| Battery Description |
| Standard Type: |
Li-ion battery Li1104c 11.1Vdc 4000mAh X2 |
| Standard Capacity: |
Up to 10 hours ECG monitoring or 60 full-energy discharges. |
| Low Battery Indicator: |
Flashing low battery icon on display and Bat Indicator on the panel of defi-monitor be on |
| Recharge Rate: |
80% in 3 hours. 100% in 5 hours. |
| Built-in Recorder |
| Paper: |
50mm thermal. |
| Printing Method: |
High-resolution, thermal print head. |
| Record date, time ,sweep speed, waveform gain, heart rate, energy (joule) and ECG waveform |
| Print mode |
HR exceed limits alarm recording |
| Auto timed recording |
| Real-Time Recording |
| Defibrillation discharge trigger recording |
| Power |
| AC-DC dual-use: |
Use the batteries,AC100V~240V,50/60 Hz. |
| Vehicle Voltage(Option): |
DC 12V |
| Main unit: |
| Portable |
| Dimensions: |
L320xW205xH410 |
| Weight: |
7.5kg |
| Pulse Oximeter (Optional) |
| Display: |
Spo2 value,waveform, Pulse |
| Probe: |
Standard adult finger clip |
| Range: |
0-99% for adult, pediatric and neonatal patients |
| AED (Optional) |
| Voice prompts and screen prompts. |
| Energy Programmable Sequence: |
200,300,360 Joules; |
| 150,150,200 Joules; |
| 50, 75, 100 Joules; |
| 30, 30, 50 Joules; |
| Discharge Control: |
Control on device front panel. |